CVE-2021-37160
- EPSS 2.03%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:45
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) du...
CVE-2021-37161
- EPSS 7.57%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:45
A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7. A buffer overflow allows an attacker to overwrite an inter...
CVE-2021-37162
- EPSS 7.57%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:45
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading ...
CVE-2021-37163
- EPSS 0.58%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:45
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded.
CVE-2021-37164
- EPSS 0.8%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:46
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An of...
CVE-2021-37166
- EPSS 0.87%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:46
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a T...
- EPSS 0.58%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:46
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the ...
CVE-2021-37165
- EPSS 5%
- Veröffentlicht 02.08.2021 11:15:11
- Zuletzt bearbeitet 21.11.2024 06:14:46
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMs...