9.8

CVE-2021-37164

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 3.4% 0.873
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 9.8 3.9 5.9
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvd@nist.gov 7.5 10 6.4
AV:N/AC:L/Au:N/C:P/I:P/A:P
CWE-787 Out-of-bounds Write

The product writes data past the end, or before the beginning, of the intended buffer.

https://www.armis.com/PwnedPiper
Broken Link
https://www.swisslog-healthcare.com
Product
https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20
https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37164-bulletin---off-by-three-stack-overflow-in-tcptxthread.pdf?rev=daf615075c71484c8059c906872a51e6&hash=1FCC1A5D921E231D71E6B95A9AA8B741
Vendor Advisory