CVE-2021-37166
- EPSS 0.87%
- Veröffentlicht 02.08.2021 13:15:07
- Zuletzt bearbeitet 21.11.2024 06:14:46
- Quelle cve@mitre.org
- CVE-Watchlists
- Unerledigt
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Swisslog-healthcare ≫ Hmi-3 Control Panel Firmware Version < 7.2.5.7
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.87% | 0.744 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 7.5 | 3.9 | 3.6 |
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
|
| nvd@nist.gov | 7.8 | 10 | 6.9 |
AV:N/AC:L/Au:N/C:N/I:N/A:C
|
CWE-120 Buffer Copy without Checking Size of Input ('Classic Buffer Overflow')
The product copies an input buffer to an output buffer without verifying that the size of the input buffer is less than the size of the output buffer, leading to a buffer overflow.