- EPSS 0.18%
- Veröffentlicht 25.03.2019 18:29:00
- Zuletzt bearbeitet 21.11.2024 02:30:08
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes...
- EPSS 0.25%
- Veröffentlicht 25.03.2019 17:29:00
- Zuletzt bearbeitet 21.11.2024 02:30:08
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP...
- EPSS 0.53%
- Veröffentlicht 25.03.2019 17:29:00
- Zuletzt bearbeitet 21.11.2024 02:30:08
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user ...
CVE-2015-3952
- EPSS 0.12%
- Veröffentlicht 25.03.2019 16:29:00
- Zuletzt bearbeitet 21.11.2024 02:30:07
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP...