7.5

CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.12% 0.278
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 7.5 3.9 3.6
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
nvd@nist.gov 5 10 2.9
AV:N/AC:L/Au:N/C:P/I:N/A:N
CWE-200 Exposure of Sensitive Information to an Unauthorized Actor

The product exposes sensitive information to an actor that is not explicitly authorized to have access to that information.

CWE-312 Cleartext Storage of Sensitive Information

The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.