CVE-2015-3958
- EPSS 1.99%
- Veröffentlicht 06.07.2015 19:59:04
- Zuletzt bearbeitet 12.04.2025 10:46:40
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
CVE-2015-3957
- EPSS 0.06%
- Veröffentlicht 06.07.2015 19:59:03
- Zuletzt bearbeitet 12.04.2025 10:46:40
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
- EPSS 14.45%
- Veröffentlicht 06.07.2015 19:59:02
- Zuletzt bearbeitet 12.04.2025 10:46:40
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
- EPSS 0.74%
- Veröffentlicht 06.07.2015 19:59:01
- Zuletzt bearbeitet 12.04.2025 10:46:40
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
CVE-2014-5406
- EPSS 0.55%
- Veröffentlicht 06.07.2015 19:59:00
- Zuletzt bearbeitet 03.11.2025 19:15:38
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data vi...
- EPSS 17.74%
- Veröffentlicht 29.04.2015 23:59:00
- Zuletzt bearbeitet 12.04.2025 10:46:40
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.