8.4

CVE-2026-17583

Medienbericht

Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check

The affected

Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.14% 0.034
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
DHS.gov 8.3 0 0
CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:H/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
DHS.gov 8.4 2.5 5.9
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CWE-353 Missing Support for Integrity Check

The product uses a transmission protocol that does not include a mechanism for verifying the integrity of the data during transmission, such as a checksum.

Für Zugriff zu Vulnerability Intelligence ist ein VulnDex Zugang erforderlich.
VulnDex Intel
Media Report
10.08.2026 18:54
https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf
Patch
Vendor Advisory
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01
Patch
Third Party Advisory
US Government Resource
https://www.cve.org/CVERecord?id=CVE-2026-17583
Third Party Advisory