CVE-2023-25931

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app.  Changes still cannot be made outside of the established therapy parameters of the programmer.  For unauthorized access to occur, an individual would need physical access to the Smart Programmer. 

Daten sind bereitgestellt durch National Vulnerability Database (NVD)
MedtronicInterstim X Clinician Versiona51300
MedtronicMicro Clinician Versiona51200
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.26% 0.489
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 6.8 0.9 5.9
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
security@medtronic.com 6.4 0.9 5.5
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
CWE-287 Improper Authentication

When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.

CWE-620 Unverified Password Change

When setting a new password for a user, the product does not require knowledge of the original password, or using another form of authentication.