8.8

CVE-2021-33886

Exploit
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Bbraun ≫ Spacecom2 Version < 012u000062
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.83% 0.526
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
NIST 8.8 2.8 5.9
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
NIST 5.8 6.5 6.4
AV:A/AC:L/Au:N/C:P/I:P/A:P
MITRE 8.1 2.8 5.2
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
CWE-134 Use of Externally-Controlled Format String

The product uses a function that accepts a format string as an argument, but the format string originates from an external source.

https://www.bbraunusa.com/en.htm
Broken Link
https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/
Third Party Advisory
Exploit