8.6

CVE-2021-33882

A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
BbraunSpacecom2 Version < 012u000062
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.31% 0.54
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 8.6 3.9 4
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N
nvd@nist.gov 5 10 2.9
AV:N/AC:L/Au:N/C:N/I:P/A:N
cve@mitre.org 6.8 2.2 4
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:N
CWE-306 Missing Authentication for Critical Function

The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.