9.9

CVE-2020-6965

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.3% 0.528
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 9.9 3.1 6
CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
nvd@nist.gov 6.5 8 6.4
AV:N/AC:L/Au:S/C:P/I:P/A:P
CWE-20 Improper Input Validation

The product receives input or data, but it does not validate or incorrectly validates that the input has the properties that are required to process the data safely and correctly.

CWE-434 Unrestricted Upload of File with Dangerous Type

The product allows the upload or transfer of dangerous file types that are automatically processed within its environment.