8.6
CVE-2020-6964
- EPSS 1.36%
- Veröffentlicht 24.01.2020 17:15:13
- Zuletzt bearbeitet 21.11.2024 05:36:23
- Erkennungen
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Gehealthcare ≫ Apexpro Telemetry Server Firmware Version <= 4.2
Gehealthcare ≫ Carescape Central Station Mai700 Firmware Version 1.0
Gehealthcare ≫ Carescape Central Station Mai700 Firmware Version 2.0
Gehealthcare ≫ Carescape Central Station Mas700 Firmware Version 1.0
Gehealthcare ≫ Carescape Central Station Mas700 Firmware Version 2.0
Gehealthcare ≫ Clinical Information Center Mp100d Firmware Version 4.0
Gehealthcare ≫ Clinical Information Center Mp100d Firmware Version 5.0
Gehealthcare ≫ Clinical Information Center Mp100r Firmware Version 4.0
Gehealthcare ≫ Clinical Information Center Mp100r Firmware Version 5.0
Gehealthcare ≫ Carescape Telemetry Server Mp100r Firmware Version <= 4.2
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 1.36% | 0.681 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| NIST | 8.6 | 3.9 | 4 |
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N
|
| NIST | 5 | 10 | 2.9 |
AV:N/AC:L/Au:N/C:N/I:P/A:N
|
CWE-306 Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
https://www.us-cert.gov/ics/advisories/icsma-20-023-01
https://www3.gehealthcare.com/~/media/downloads/us/support/site-planning/site-readiness/gehc-gateway_project_implementation_guide_pdf.pdf