10

CVE-2020-25187

Medtronic MyCareLink Smart 25000 is 

 vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 1.27% 0.788
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 9.8 3.9 5.9
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvd@nist.gov 10 10 10
AV:N/AC:L/Au:N/C:C/I:C/A:C
ics-cert@hq.dhs.gov 8.8 2.1 6
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
CWE-122 Heap-based Buffer Overflow

A heap overflow condition is a buffer overflow, where the buffer that can be overwritten is allocated in the heap portion of memory, generally meaning that the buffer was allocated using a routine such as malloc().

CWE-787 Out-of-bounds Write

The product writes data past the end, or before the beginning, of the intended buffer.