6.5

CVE-2020-16224

In Patient Information Center iX (PICiX) Versions C.02, C.03, the 
software parses a formatted message or structure but does not handle or 
incorrectly handles a length field that is inconsistent with the actual 
length of the associated data, causing the application on the 
surveillance station to restart.



Data is provided by the National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Type Source Score Percentile
EPSS FIRST.org 0.07% 0.186
CVSS Metriken
Source Base Score Exploit Score Impact Score Vector string
nvd@nist.gov 6.5 2.8 3.6
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
nvd@nist.gov 3.3 6.5 2.9
AV:A/AC:L/Au:N/C:N/I:N/A:P
CWE-130 Improper Handling of Length Parameter Inconsistency

The product parses a formatted message or structure, but it does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data.