5.8

CVE-2020-16214

Philips Patient Monitoring Devices Improper Neutralization of Formula Elements in a CSV File

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the 
software saves user-provided information into a comma-separated value 
(CSV) file, but it does not neutralize or incorrectly neutralizes 
special elements that could be interpreted as a command when the file is
 opened by spreadsheet software.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.14% 0.307
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 5 1.8 2.7
CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
nvd@nist.gov 5.8 8.6 4.9
AV:N/AC:M/Au:N/C:P/I:P/A:N
CWE-1236 Improper Neutralization of Formula Elements in a CSV File

The product saves user-provided information into a Comma-Separated Value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by a spreadsheet product.