6.5

CVE-2020-15486

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
DrtrustElectrocardiogram Pen Firmware Version2.00.08
   DrtrustElectrocardiogram Pen Version2.00.08
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.51% 0.392
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 6.5 2.8 3.6
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
nvd@nist.gov 3.3 6.5 2.9
AV:A/AC:L/Au:N/C:P/I:N/A:N
Es wurden noch keine Informationen zu CWE veröffentlicht.
https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-in-dr-trust-ecg-or-ekg-pen
Third Party Advisory