9.8
CVE-2020-12043
- EPSS 2.1%
- Veröffentlicht 29.06.2020 14:15:11
- Zuletzt bearbeitet 21.11.2024 04:59:10
- Erkennungen
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Baxter ≫ Sigma Spectrum Infusion System Firmware Version 8.0
Baxter ≫ Sigma Spectrum Infusion System Version -
Baxter ≫ Wireless Battery Module Version 17
Baxter ≫ Wireless Battery Module Version 20d29
Baxter ≫ Wireless Battery Module Version 20d30
Baxter ≫ Wireless Battery Module Version 20d31
Baxter ≫ Wireless Battery Module Version 22d24
Baxter ≫ Wireless Battery Module Version 17
Baxter ≫ Wireless Battery Module Version 20d29
Baxter ≫ Wireless Battery Module Version 20d30
Baxter ≫ Wireless Battery Module Version 20d31
Baxter ≫ Wireless Battery Module Version 22d24
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 2.1% | 0.802 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| NIST | 9.8 | 3.9 | 5.9 |
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
|
| NIST | 7.5 | 10 | 6.4 |
AV:N/AC:L/Au:N/C:P/I:P/A:P
|
CWE-672 Operation on a Resource after Expiration or Release
The product uses, accesses, or otherwise operates on a resource after that resource has been expired, released, or revoked.
https://www.us-cert.gov/ics/advisories/icsma-20-170-04