9.8

CVE-2020-12040

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.2% 0.422
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 9.8 3.9 5.9
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvd@nist.gov 5 10 2.9
AV:N/AC:L/Au:N/C:P/I:N/A:N
CWE-319 Cleartext Transmission of Sensitive Information

The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.