Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Baxter ≫ Sigma Spectrum Infusion System Firmware Version >= 6.0 <= 6.05
Baxter ≫ Sigma Spectrum Infusion System Firmware Version8.0
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.06% | 0.141 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 2.4 | 0.9 | 1.4 |
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
|
| nvd@nist.gov | 2.1 | 3.9 | 2.9 |
AV:L/AC:L/Au:N/C:P/I:N/A:N
|
CWE-259 Use of Hard-coded Password
The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.
CWE-798 Use of Hard-coded Credentials
The product contains hard-coded credentials, such as a password or cryptographic key.