Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Baxter ≫ Prismax Firmware Version < 3.0
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.05% | 0.113 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 4.9 | 0.7 | 4.2 |
CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L
|
| nvd@nist.gov | 3.6 | 3.9 | 4.9 |
AV:L/AC:L/Au:N/C:N/I:P/A:P
|
CWE-287 Improper Authentication
When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.
CWE-798 Use of Hard-coded Credentials
The product contains hard-coded credentials, such as a password or cryptographic key.