CVE-2020-10627

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.09% 0.252
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 8.1 2.8 5.2
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
nvd@nist.gov 4.8 6.5 4.9
AV:A/AC:L/Au:N/C:P/I:P/A:N
ics-cert@hq.dhs.gov 7.3 2.5 4.7
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L
CWE-284 Improper Access Control

The product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.