9.3

CVE-2019-6538

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
MedtronicCarelink Monitor Firmware Version2490c
   MedtronicCarelink Monitor Version-
MedtronicCarelink 2090 Firmware Version-
   MedtronicCarelink 2090 Version-
MedtronicAmplia Crt-d Firmware Version-
   MedtronicAmplia Crt-d Version-
MedtronicClaria Crt-d Firmware Version-
   MedtronicClaria Crt-d Version-
MedtronicCompia Crt-d Firmware Version-
   MedtronicCompia Crt-d Version-
MedtronicConcerto Crt-d Firmware Version-
   MedtronicConcerto Crt-d Version-
MedtronicConsulta Crt-d Firmware Version-
   MedtronicConsulta Crt-d Version-
MedtronicEvera Icd Firmware Version-
   MedtronicEvera Icd Version-
MedtronicMirro Icd Firmware Version-
   MedtronicMirro Icd Version-
MedtronicNayamed Nd Icd Firmware Version-
   MedtronicNayamed Nd Icd Version-
MedtronicPrimo Icd Firmware Version-
   MedtronicPrimo Icd Version-
MedtronicSecura Icd Firmware Version-
   MedtronicSecura Icd Version-
MedtronicVirtuoso Icd Firmware Version-
   MedtronicVirtuoso Icd Version-
MedtronicVisia Af Icd Firmware Version-
   MedtronicVisia Af Icd Version-
MedtronicViva Crt-d Firmware Version-
   MedtronicViva Crt-d Version-
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.22% 0.445
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 6.5 2.8 3.6
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
nvd@nist.gov 3.3 6.5 2.9
AV:A/AC:L/Au:N/C:N/I:P/A:N
ics-cert@hq.dhs.gov 9.3 2.8 5.8
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
CWE-284 Improper Access Control

The product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.

CWE-306 Missing Authentication for Critical Function

The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

CWE-862 Missing Authorization

The product does not perform an authorization check when an actor attempts to access a resource or perform an action.