6.5

CVE-2019-6538

Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Medtronic ≫ Mycarelink Monitor Firmware Version 24950
   Medtronic ≫ Mycarelink Monitor Version -
Medtronic ≫ Mycarelink Monitor Firmware Version 24952
   Medtronic ≫ Mycarelink Monitor Version -
Medtronic ≫ Carelink Monitor Firmware Version 2490c
   Medtronic ≫ Carelink Monitor Version -
Medtronic ≫ Carelink 2090 Firmware Version -
   Medtronic ≫ Carelink 2090 Version -
Medtronic ≫ Amplia Crt-d Firmware Version -
   Medtronic ≫ Amplia Crt-d Version -
Medtronic ≫ Claria Crt-d Firmware Version -
   Medtronic ≫ Claria Crt-d Version -
Medtronic ≫ Compia Crt-d Firmware Version -
   Medtronic ≫ Compia Crt-d Version -
Medtronic ≫ Concerto Crt-d Firmware Version -
   Medtronic ≫ Concerto Crt-d Version -
Medtronic ≫ Consulta Crt-d Firmware Version -
   Medtronic ≫ Consulta Crt-d Version -
Medtronic ≫ Evera Icd Firmware Version -
   Medtronic ≫ Evera Icd Version -
Medtronic ≫ Mirro Icd Firmware Version -
   Medtronic ≫ Mirro Icd Version -
Medtronic ≫ Nayamed Nd Icd Firmware Version -
   Medtronic ≫ Nayamed Nd Icd Version -
Medtronic ≫ Primo Icd Firmware Version -
   Medtronic ≫ Primo Icd Version -
Medtronic ≫ Secura Icd Firmware Version -
   Medtronic ≫ Secura Icd Version -
Medtronic ≫ Virtuoso Icd Firmware Version -
   Medtronic ≫ Virtuoso Icd Version -
Medtronic ≫ Virtuoso Ii Icd Firmware Version -
   Medtronic ≫ Virtuoso Ii Icd Version -
Medtronic ≫ Visia Af Icd Firmware Version -
   Medtronic ≫ Visia Af Icd Version -
Medtronic ≫ Viva Crt-d Firmware Version -
   Medtronic ≫ Viva Crt-d Version -
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.9% 0.555
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
NIST 6.5 2.8 3.6
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
NIST 3.3 6.5 2.9
AV:A/AC:L/Au:N/C:N/I:P/A:N
DHS.gov 9.3 2.8 5.8
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
CWE-284 Improper Access Control

The product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.

CWE-306 Missing Authentication for Critical Function

The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.

CWE-862 Missing Authorization

The product does not perform an authorization check when an actor attempts to access a resource or perform an action.

http://www.securityfocus.com/bid/107544
Third Party Advisory
VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01
Third Party Advisory
US Government Resource
Mitigation