9.3
CVE-2019-6538
- EPSS 0.84%
- Veröffentlicht 25.03.2019 22:29:00
- Zuletzt bearbeitet 22.05.2025 20:15:20
- Quelle ics-cert@hq.dhs.gov
- CVE-Watchlists
- Unerledigt
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Medtronic ≫ Mycarelink Monitor Firmware Version24950
Medtronic ≫ Mycarelink Monitor Firmware Version24952
Medtronic ≫ Carelink Monitor Firmware Version2490c
Medtronic ≫ Carelink 2090 Firmware Version-
Medtronic ≫ Amplia Crt-d Firmware Version-
Medtronic ≫ Claria Crt-d Firmware Version-
Medtronic ≫ Compia Crt-d Firmware Version-
Medtronic ≫ Concerto Crt-d Firmware Version-
Medtronic ≫ Concerto Ii Crt-d Firmware Version-
Medtronic ≫ Consulta Crt-d Firmware Version-
Medtronic ≫ Evera Icd Firmware Version-
Medtronic ≫ Maximo Ii Crt-d And Lcd Firmware Version-
Medtronic ≫ Mirro Icd Firmware Version-
Medtronic ≫ Nayamed Nd Icd Firmware Version-
Medtronic ≫ Primo Icd Firmware Version-
Medtronic ≫ Protecta Icd And Crt-d Firmware Version-
Medtronic ≫ Secura Icd Firmware Version-
Medtronic ≫ Virtuoso Icd Firmware Version-
Medtronic ≫ Virtuoso Ii Icd Firmware Version-
Medtronic ≫ Visia Af Icd Firmware Version-
Medtronic ≫ Viva Crt-d Firmware Version-
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.84% | 0.531 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 6.5 | 2.8 | 3.6 |
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
|
| nvd@nist.gov | 3.3 | 6.5 | 2.9 |
AV:A/AC:L/Au:N/C:N/I:P/A:N
|
| ics-cert@hq.dhs.gov | 9.3 | 2.8 | 5.8 |
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
|
CWE-284 Improper Access Control
The product does not restrict or incorrectly restricts access to a resource from an unauthorized actor.
CWE-306 Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.
CWE-862 Missing Authorization
The product does not perform an authorization check when an actor attempts to access a resource or perform an action.
http://www.securityfocus.com/bid/107544
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01