4.8

CVE-2017-14018

An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.14% 0.352
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 4.8 0.5 4.2
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L
nvd@nist.gov 3.3 3.4 4.9
AV:L/AC:M/Au:N/C:N/I:P/A:P
CWE-287 Improper Authentication

When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.