6.8

CVE-2017-12725

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.85% 0.534
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 5.6 2.2 3.4
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
nvd@nist.gov 6.8 8.6 6.4
AV:N/AC:M/Au:N/C:P/I:P/A:P
CWE-798 Use of Hard-coded Credentials

The product contains hard-coded credentials, such as a password or cryptographic key.

http://www.securityfocus.com/bid/100665
Third Party Advisory
VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A
Third Party Advisory
US Government Resource