6.5
CVE-2017-12716
- EPSS 0.04%
- Published 25.04.2018 13:29:00
- Last modified 21.11.2024 03:10:05
- Source ics-cert@hq.dhs.gov
- Teams watchlist Login
- Open Login
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Data is provided by the National Vulnerability Database (NVD)
Abbott ≫ Accent Firmware Version < f0b.0e.7e
Abbott ≫ Anthem Firmware Version < f0b.0e.7e
Abbott ≫ Accent Mri Firmware Version < f10.08.6c
Abbott ≫ Accent St Firmware Version < f10.08.6c
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
Type | Source | Score | Percentile |
---|---|---|---|
EPSS | FIRST.org | 0.04% | 0.081 |
Source | Base Score | Exploit Score | Impact Score | Vector string |
---|---|---|---|---|
nvd@nist.gov | 6.5 | 2.8 | 3.6 |
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
|
nvd@nist.gov | 3.3 | 6.5 | 2.9 |
AV:A/AC:L/Au:N/C:P/I:N/A:N
|
CWE-311 Missing Encryption of Sensitive Data
The product does not encrypt sensitive or critical information before storage or transmission.
CWE-319 Cleartext Transmission of Sensitive Information
The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.