6.5

CVE-2017-12716

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Data is provided by the National Vulnerability Database (NVD)
AbbottAccent Firmware Version < f0b.0e.7e
   AbbottAccent Version-
AbbottAnthem Firmware Version < f0b.0e.7e
   AbbottAnthem Version-
AbbottAccent Mri Firmware Version < f10.08.6c
   AbbottAccent Mri Version-
AbbottAccent St Firmware Version < f10.08.6c
   AbbottAccent St Version-
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Type Source Score Percentile
EPSS FIRST.org 0.04% 0.081
CVSS Metriken
Source Base Score Exploit Score Impact Score Vector string
nvd@nist.gov 6.5 2.8 3.6
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
nvd@nist.gov 3.3 6.5 2.9
AV:A/AC:L/Au:N/C:P/I:N/A:N
CWE-311 Missing Encryption of Sensitive Data

The product does not encrypt sensitive or critical information before storage or transmission.

CWE-319 Cleartext Transmission of Sensitive Information

The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.