9.9
CVE-2016-8355
- EPSS 0.39%
- Veröffentlicht 13.02.2017 22:59:00
- Zuletzt bearbeitet 20.04.2025 01:37:25
- Quelle ics-cert@hq.dhs.gov
- CVE-Watchlists
- Unerledigt
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Smiths-medical ≫ Cadd-solis Medication Safety Software Version1.0
Smiths-medical ≫ Cadd-solis Medication Safety Software Version2.0
Smiths-medical ≫ Cadd-solis Medication Safety Software Version3.0
Smiths-medical ≫ Cadd-solis Medication Safety Software Version3.1
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.39% | 0.573 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 9.9 | 3.1 | 6 |
CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
|
| nvd@nist.gov | 9 | 8 | 10 |
AV:N/AC:L/Au:S/C:C/I:C/A:C
|
CWE-306 Missing Authentication for Critical Function
The product does not perform any authentication for functionality that requires a provable user identity or consumes a significant amount of resources.