7.5

CVE-2015-1012

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.77% 0.508
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 7.5 3.9 3.6
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
nvd@nist.gov 5 10 2.9
AV:N/AC:L/Au:N/C:P/I:N/A:N
CWE-200 Exposure of Sensitive Information to an Unauthorized Actor

The product exposes sensitive information to an actor that is not explicitly authorized to have access to that information.

CWE-312 Cleartext Storage of Sensitive Information

The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.

https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
Third Party Advisory
US Government Resource
Mitigation