6.8

CVE-2014-5431

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.05% 0.133
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
nvd@nist.gov 6.8 0.9 5.9
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvd@nist.gov 4.6 3.9 6.4
AV:L/AC:L/Au:N/C:P/I:P/A:P
CWE-259 Use of Hard-coded Password

The product contains a hard-coded password, which it uses for its own inbound authentication or for outbound communication to external components.

CWE-798 Use of Hard-coded Credentials

The product contains hard-coded credentials, such as a password or cryptographic key.