8.2

CVE-2025-0683

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text 
patient data to a hard-coded public IP address when a patient is hooked 
up to the monitor. This could lead to a leakage of confidential patient 
data to any device with that IP address or an attacker in a 
machine-in-the-middle scenario.
Daten sind bereitgestellt durch das CVE Programm von einer CVE Numbering Authority (CNA) (Unstrukturiert).
HerstellerContec Health
Produkt CMS8000 Patient Monitor
Default Statusunaffected
Version All versions
Status affected
VulnDex Vulnerability Enrichment
Diese Information steht angemeldeten Benutzern zur Verfügung. Login Login
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 0.77% 0.506
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
ics-cert@hq.dhs.gov 8.2 0 0
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
ics-cert@hq.dhs.gov 5.9 2.2 3.6
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CWE-359 Exposure of Private Personal Information to an Unauthorized Actor

The product does not properly prevent a person's private, personal information from being accessed by actors who either (1) are not explicitly authorized to access the information or (2) do not have the implicit consent of the person about whom the information is collected.

https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor