5.3
CVE-2022-43557
- EPSS 0.11%
- Veröffentlicht 05.12.2022 22:15:11
- Zuletzt bearbeitet 21.11.2024 07:26:47
- Quelle cybersecurity@bd.com
- CVE-Watchlists
- Unerledigt
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
Bd ≫ Bodyguard 999-603 Firmware Version-
Bd ≫ Bodyguard Duo 999-903 Firmware Version-
Bd ≫ Bodyguard Epidural 999-683 Firmware Version-
Bd ≫ Bodyguard Pain Manager 999-803 Firmware Version-
Bd ≫ Bodyguard T 999-103 Firmware Version-
Bd ≫ Bodyguard 323 Colorvision Firmware Version-
Bd ≫ Bodyguard 121 Twins Firmware Version-
| Typ | Quelle | Score | Percentile |
|---|---|---|---|
| EPSS | FIRST.org | 0.11% | 0.294 |
| Quelle | Base Score | Exploit Score | Impact Score | Vector String |
|---|---|---|---|---|
| nvd@nist.gov | 5.3 | 0.5 | 4.7 |
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H
|
| cybersecurity@bd.com | 5.3 | 0.5 | 4.7 |
CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H
|
CWE-1299 Missing Protection Mechanism for Alternate Hardware Interface
The lack of protections on alternate paths to access control-protected assets (such as unprotected shadow registers and other external facing unguarded interfaces) allows an attacker to bypass existing protections to the asset that are only performed against the primary path.
CWE-287 Improper Authentication
When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.