9

CVE-2020-25156

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
Daten sind bereitgestellt durch National Vulnerability Database (NVD)
BbraunDatamodule Compactplus Versiona10
   BbraunDatamodule Compactplus Version-
BbraunDatamodule Compactplus Versiona11
   BbraunDatamodule Compactplus Version-
BbraunSpacecom Version <= l81
   BbraunSpacecom Version-
Zu dieser CVE wurde keine Warnung gefunden.
EPSS Metriken
Typ Quelle Score Percentile
EPSS FIRST.org 1.17% 0.645
CVSS Metriken
Quelle Base Score Exploit Score Impact Score Vector String
NIST 7.2 1.2 5.9
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
NIST 9 8 10
AV:N/AC:L/Au:S/C:C/I:C/A:C
DHS.gov 7.2 1.2 5.9
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
CWE-489 Active Debug Code

The product is released with debugging code still enabled or active.

https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Broken Link
https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02
Third Party Advisory