6.5
CVE-2017-12714
- EPSS 0.22%
- Published 25.04.2018 13:29:00
- Last modified 21.11.2024 03:10:04
- Source ics-cert@hq.dhs.gov
- Teams watchlist Login
- Open Login
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Data is provided by the National Vulnerability Database (NVD)
Abbott ≫ Accent Firmware Version < f0b.0e.7e
Abbott ≫ Anthem Firmware Version < f0b.0e.7e
Abbott ≫ Accent Mri Firmware Version < f10.08.6c
Abbott ≫ Accent St Firmware Version < f10.08.6c
Abbott ≫ Assurity Firmware Version < f14.07.80
Abbott ≫ Allure Firmware Version < f14.07.80
Abbott ≫ Assurity Mri Firmware Version < f17.01.49
Zu dieser CVE wurde keine CISA KEV oder CERT.AT-Warnung gefunden.
Type | Source | Score | Percentile |
---|---|---|---|
EPSS | FIRST.org | 0.22% | 0.411 |
Source | Base Score | Exploit Score | Impact Score | Vector string |
---|---|---|---|---|
nvd@nist.gov | 6.5 | 2.8 | 3.6 |
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
|
nvd@nist.gov | 6.1 | 6.5 | 6.9 |
AV:A/AC:L/Au:N/C:N/I:N/A:C
|
CWE-920 Improper Restriction of Power Consumption
The product operates in an environment in which power is a limited resource that cannot be automatically replenished, but the product does not properly restrict the amount of power that its operation consumes.